RESISTANCE HiTESTER
RM3543
Resolution and Strong Noise Immunity for Integration
into Automated Production Lines
Hioki DC resistance meters measure a broad range of resistance values at a high level of precision. The RM3543 is designed to be integrated into current sensing resistor taping machines, deliver high-accuracy measurement of super-low-resistance levels, and are advanced enough to measure 0.1 mÎĐ shunts with room to spare. |
Model No. (Order Code)
Test fixtures are not supplied with the unit. |
---|
Key Features âĒ Advanced enough to measure 0.1 mÎĐ shunts with room to spare at Âą0.16% accuracy & 0.01ΞÎĐ resolution performance |
---|
A resistance tester that can measure super-low resistance values and shunt resistors
The Resistance Meter RM3543 can measure low shunt resistance values and other DC |
|
Reduce the effects of thermoelectromotive force with the offset voltage correction
Thermoelectromotive force refers to the potential difference that occurs at junctions between |
|
Use the scaling function to correct measurements as if the target component were mounted
The RM3543âs scaling function can correct for the difference in resistance values that are |
|
Ensure reliable contact quickly with the contact improvement function
The RM3543 improves contact with measurement specimens by eliminating oxide films and dirt |
|
Using probe short detection functionality to ensure the reliability of 4-terminal measurement
The presence of any foreign material between the POT and CUR measurement probes can interfere with |
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement method | Four-terminal, constant-current DC |
---|---|
Resistance range | 10 mÎĐ (Max. 12.00000 mÎĐ, 0.01 ΞÎĐ resolution) to 1000 ÎĐ range (Max. 1200.000 ÎĐ, 1 mÎĐ resolution), 7 steps |
Display | Monochrome graphic LCD 240 Ã 64 dot, white LED backlight |
Measurement accuracy | [at 10 mÎĐ range, with SLOW mode, average 16 times settings] Âą0.060 % rdg. Âą0.001 % f.s. |
Testing current | [at 10 mÎĐ range] 1 A DC to [at 1000 ÎĐ range] 1 mA DC |
Open-terminal voltage | 20 V DC max. Note: Voltage when not measuring is 20 mV or less, with current mode set at PULSE and Contact Improver Setting set at OFF/PULSE (measured with a voltmeter having 10 MÎĐ) |
Sampling rate | FAST, MEDIUM, SLOW, 3 settings |
Integration time | [at 10 mÎĐ range, default value] FAST 2.0 ms, MED 5.0 ms, SLOW 1 PLC, Setting range: 0.1 ms to 100.0 ms, or 1 to 5 PLC at 50 Hz, 1 to 6 PLC at 60 Hz Note: PLC = one power line cycle (mains wave-form period) |
Other functions | Comparator (compare setting value with measurement value), Delay, OVC (offset voltage compensation), Average, Measurement fault detection, Probe short-circuit detection, Improve contact, Current mode setting (A pulse application function that applies current only during measurement), Auto-memory, Statistical calculations, Settings monitor (when using two instruments, a difference in settings causes warning notification), Retry, Trigger function. etc,. |
Interfaces | External I/O, RS-232C, Printer, GP-IB (Model RM3543-01) |
External I/O | Trigger, Hold input, Comparator output, Settings monitor terminal, Service power output +5V, +12V, etc. |
Power supply | 100 to 240 V AC, 50/60 Hz, 40 VA max. |
Dimensions and mass | 260 mm (10.24 in) W Ã 88 mm (3.46 in) H Ã 300 mm (11.81 in) D, 3.0 kg (105.8 oz) |
Accessories | Power cord Ã1, EXT. I/O male connector Ã1, Instruction manual Ã1, Operation guide Ã1 |
Catalog: RESISTANCE HiTESTER RM3543 | Download PDFÂ [1MB] | English |