RESISTANCE HiTESTER
RM3542
Systems
Hioki DC resistance meters measure a broad range of resistance values at a high level of precision. The RM3542 deliver high speed, high precision, and high reliability, indispensable for the shipping inspection of devices such as resistors and EMI components. The resistance meter also supports chip resistor sampling inspections, helping resistor manufacturers improve productivity. |
Model No. (Order Code)
Test fixtures are not supplied with the unit. |
---|
Key Features âĒ High speed and accuracy maximize productivity in automated systems |
---|
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Resistance range | [at Low Power OFF] 100 mÎĐ range (Max. 120.0000 mÎĐ, 0.1 ΞÎĐ resolution) to 100 MÎĐ range (Max. 120.0000 MÎĐ, 100 ÎĐ resolution), 10 steps [at Low Power ON] 1000 mÎĐ range (Max. 1200.000 mÎĐ, 1 ΞÎĐ resolution) to 1000 ÎĐ range (Max. 1200.000 ÎĐ, 1 mÎĐ resolution), 4 steps |
---|---|
Display | Monochrome graphic LCD 240 Ã 64 dot, white LED backlight |
Measurement accuracy | [with SLOW mode, at 100 mÎĐ range] Âą0.015 % rdg. Âą0.002 % f.s. [with SLOW mode, at 1000 ÎĐ range] Âą0.006 % rdg. Âą0.001 % f.s. (the best case) |
Testing current | [at 100 mÎĐ range] 100 mA DC to [at 100 MÎĐ range] 100 nA DC |
Open-terminal voltage | 20 V DC max. |
Sampling rate | FAST, MEDIUM, SLOW, 3 settings |
Measurement times | [at 100 ÎĐ /1000 ÎĐ ranges, with Low Power OFF] FAST: 0.9 ms, MED: 3.6 ms, SLOW: 17 ms (minimum time) |
Integration time | 0.1 ms to 100.0 ms, or 1 to 5 PLC at 50 Hz, 1 to 6 PLC at 60 Hz Note: PLC = one power line cycle (mains wave-form period) |
Other functions | Comparator (compare setting value with measurement value), Delay (set to allow for mechanical delay of trigger input and probing, or set to allow for measurement object response), OVC (offset voltage compensation), Measurement fault detection, Probe short-circuit detection, Improve contact, Auto-memory, Statistical calculations, Settings monitor (when using two instruments, a difference in settings causes warning notification), Retry, Trigger function. etc,. |
Interfaces | RS-232C, Printer, GP-IB (Model RM3542-01) |
External I/O | Trigger, Hold input, Comparator output, Settings monitor terminal |
Power supply | 100 to 240 V AC, 50/60 Hz, 30 VA max. |
Dimensions and mass | 260 mm (10.24 in) W Ã 88 mm (3.46 in) H Ã 300 mm (11.81 in) D, 2.9 kg (102.3 oz) |
Accessories | Power cord Ã1, EXT. I/O male connector Ã1, Instruction manual Ã1, Operation guide Ã1 |
Catalog: RESISTANCE METER RM3542A, RM3542 | Download PDFÂ [2MB] | English |