RESISTANCE HiTESTER
RM3542

Measure in as Fast as 0.9 ms, Optimized for Automated
Systems

Hioki DC resistance meters measure a broad range of resistance values at a
high level of precision. The RM3542 deliver high speed, high precision,
and high reliability, indispensable for the shipping inspection of
devices such as resistors and EMI components. The resistance meter also
supports chip resistor sampling inspections, helping resistor manufacturers
improve productivity.

Model No. (Order Code)

RM3542
RM3542-01 Built in GP-IB interface
GPIBRS232CCE

Test fixtures are not supplied with the unit.
Please select an optional test fixture when ordering.

Key Features

â€Ē High speed and accuracy maximize productivity in automated systems
â€Ē Multiple checking functions ensure proper contact for reliable measurements
â€Ē Low-power resistance mode measures chip inductors and EMC suppression components
â€Ē Supports sample inspections during the manufacturing process

Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Resistance range [at Low Power OFF] 100 mÎĐ range (Max. 120.0000 mÎĐ, 0.1 ΞÎĐ resolution) to 100 MÎĐ range (Max. 120.0000 MÎĐ, 100 ÎĐ resolution), 10 steps
[at Low Power ON] 1000 mÎĐ range (Max. 1200.000 mÎĐ, 1 ΞÎĐ resolution) to 1000 ÎĐ range (Max. 1200.000 ÎĐ, 1 mÎĐ resolution), 4 steps
Display Monochrome graphic LCD 240 × 64 dot, white LED backlight
Measurement accuracy [with SLOW mode, at 100 mÎĐ range] Âą0.015 % rdg. Âą0.002 % f.s.
[with SLOW mode, at 1000 ÎĐ range] Âą0.006 % rdg. Âą0.001 % f.s. (the best case)
Testing current [at 100 mÎĐ range] 100 mA DC to [at 100 MÎĐ range] 100 nA DC
Open-terminal voltage 20 V DC max.
Sampling rate FAST, MEDIUM, SLOW, 3 settings
Measurement times [at 100 ÎĐ /1000 ÎĐ ranges, with Low Power OFF]
FAST: 0.9 ms, MED: 3.6 ms, SLOW: 17 ms (minimum time)
Integration time 0.1 ms to 100.0 ms, or 1 to 5 PLC at 50 Hz, 1 to 6 PLC at 60 Hz
Note: PLC = one power line cycle (mains wave-form period)
Other functions Comparator (compare setting value with measurement value), Delay (set to allow for mechanical delay of trigger input and probing, or set to allow for measurement object response), OVC (offset voltage compensation), Measurement fault detection, Probe short-circuit detection, Improve contact, Auto-memory, Statistical calculations, Settings monitor (when using two instruments, a difference in settings causes warning notification), Retry, Trigger function. etc,.
Interfaces RS-232C, Printer, GP-IB (Model RM3542-01)
External I/O Trigger, Hold input, Comparator output, Settings monitor terminal
Power supply 100 to 240 V AC, 50/60 Hz, 30 VA max.
Dimensions and mass 260 mm (10.24 in) W × 88 mm (3.46 in) H × 300 mm (11.81 in) D, 2.9 kg (102.3 oz)
Accessories Power cord ×1, EXT. I/O male connector ×1, Instruction manual ×1, Operation guide ×1
Catalog: RESISTANCE METER RM3542A, RM3542 Download PDF  [2MB] English