IMPEDANCE ANALYZER
IM7585
Analyzer with 0.5ms Test Speed and 0.07% Variability
Hioki LCR Meters and Impedance Analyzers range from 1mHz to 3GHz devices to suit a wide range of applications in the testing of electronic components. The IM7585 Impedance Analyzer offers a top measurement time of 0.5ms over a 1MHz to 1.3GHz frequency range and superior stability, making it ideal for R&D as well as high volume production of ferrite chip beads and chip inductors. Key Features |
Model No. (Order Code)
The instrument does not ship with a test fixture or probe. A test fixture designed specifically for use with the Impedance Analyzer is required. |
IMPEDANCE ANALYZER IM7580 seriesïžLCR Measurement Using the SMD Test Fixture IM9201 The IM7580 series encompasses five models covering measurement frequencies of 1 MHz to 3 GHz. Used in combination with the Test Fixture IM9201, which can accommodate six SMD sizes, IM7580 series instruments allow you to measure specimens easily and reliably. |
|
IMPEDANCE ANALYZER IM7580 seriesïžComparator function A comparator function lets you check whether measured values fall within a user-defined judgment range. This functionality is ideal for use in generating PASS/FAIL judgments for specimens. |
|
IMPEDANCE ANALYZER IM7580 seriesïžContact check function The instrumentsâ contact check function lets you check contact between the specimen and measurement terminals to detect faulty contact or verify good contact. |
|
IMPEDANCE ANALYZER IM7580 seriesïžHow to connect the test head Connect the cable of the test head to the impedance analyzer. Tighten the nut until the handle of the torque wrench bends slightly. Do not over tighten. |
|
Application information Download the sample application and view the Communications Command User Manual. |
|
Accuracy calculation information Calculate accuracy by simply entering a numerical value. |
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement modes | LCR mode, Analyzer mode (sweeps with measurement frequency and measurement level), Continuous measurement mode |
---|---|
Measurement parameters | Z, Y, Îļ, Rs (ESR), Rp, X, G, B, Cs, Cp, Ls, Lp, D (tanÎī), Q |
Measurable range | 100 mÎĐ to 5 kÎĐ |
Display range | Z: 0.00 m to 9.99999 GÎĐ / Rs, Rp, X: Âą (0.00 m to 9.99999 GÎĐ) Ls, Lp: Âą (0.00000 n to 9.99999 GH) / Q: Âą (0.00 to 9999.99) Îļ: Âą (0.000° to 180.000°), Cs, Cp: Âą (0.00000 p to 9.99999 GF) D: Âą (0.00000 to 9.99999), Y: (0.000 n to 9.99999 GS) G, B: Âą (0.000 n to 9.99999 GS), Î%: Âą (0.000 % to 999.999 %) |
Basic accuracy | Z: Âą0.65 % rdg. Îļ: Âą0.38° |
Measurement frequency | 1 MHz to 1.3 GHz (100 kHz setting resolution) |
Measurement signal level | Power: -40.0 dBm to +1.0 dBm Voltage: 4 mV to 502 mVrms Current: 0.09 mA to 10.04 mArms |
Output impedance | 50 ÎĐ (at 10 MHz) |
Display | 8.4-inch color TFT with touch screen |
Measurement speeds | FAST: 0.5 ms (Analog measurement time, typical value) |
Functions | Contact check, Comparator, BIN measurement (classification), Panel loading/saving, Memory function, Equivalent circuit analysis, Correlation compensation |
Interfaces | EXT I/O (Handler), USB communication, USB memory, LAN, RS-232C (optional), GP-IB (optional) |
Power supply | 100 to 240 V AC, 50/60 Hz, 70 VA max. |
Dimensions and mass | Main unit: 215 mm (8.46 in) W Ã 200 mm (7.87 in) H Ã 348 mm (13.70 in) D, 8.0 kg (282.2 oz) Test head: 90 mm (3.54 in) W Ã 64 mm (2.52 in) H Ã 24 mm (0.94 in) D, 300 g (10.58 oz) |
Accessories | Test head Ã1, Connection cable Ã1, Instruction manual Ã1, LCR application disc (Communications user manual) Ã1, Power cord Ã1 |
Catalog: SMD TEST FIXTURE IM9201 | Download PDFÂ [1MB] | English |
Catalog: IMPEDANCE ANALYZER IM7580 Series | Download PDFÂ [3MB] | English |
For more information, click here