LOGIC PROBE
9320-01

Detect the Presence of Voltage or Relay Contact Signals
with Memory HiCorders

â€Ē 4 input channels (Not isolated)
â€Ē For ON/OFF detection of voltage/contact signals
â€Ē Select from 3 steps of input thresholds
â€Ē Response pulse width of 500 ns or more
â€Ē Compatible with Memory HiCorders featuring a miniature terminal

Model No. (Order Code)

9320-01 For the Memory HiCorder, miniature terminal type

Note: Verify that your Memory HiCorder features a miniature connection terminal before purchase

CE

Basic specifications
Note: The unit-side plug of the 9320-01 is different from the 9320

Function Detection of voltage signal or relay contact signal for High/Low state recording
Input 4 channels (common ground between unit and channels), digital/contact input, switchable (contact input can detect open-collector signals)
Input resistance: 1 MÎĐ (with digital input, 0 to +5 V)
500 kÎĐ or more (with digital input, +5 to +50V)
Pull-up resistance: 2 kÎĐ (contact input: internally pulled up to +5 V)
Digital input threshold 1.4V/ 2.5V/ 4.0V
Contact input detection resistance 1.4 V: 1.5 kÎĐ or higher (open) and 500 ÎĐ or lower (short)
2.5 V: 3.5 kÎĐ or higher (open) and 1.5 kÎĐ or lower (short)
4.0 V: 25 kÎĐ or higher (open) and 8 kÎĐ or lower (short)
Detectable pulse width 500 ns or longer
Max. allowable input 0 to +50V DC (the maximum voltage that can be applied across input pins without damage)
Dimensions and mass 62 mm (2.44 in)W × 94 mm (3.70 in)H × 20 mm (0.79 in)D, 150 g (5.3 oz), Main unit cable 1.5 m (4.92 ft), input section cable 300 mm (0.98 ft)
Accessories IC clip leads ×1, Alligator clip leads ×1, Carrying case ×1, Instruction Manual ×1