AC/DC CURRENT SENSOR
CT6876
Exclusive Academic Offer
Hioki pass-through AC/DC current sensors are best-in-class devices for use with power meters, Memory HiCorders and high performance oscilloscopes. The CT6876 is a DC to 1.5MHz/1000A sensor, ideal for testing inverter efficiency, inverter output power, reactor or transformer loss.
The sensor uses flux gate detection within the zero flux method to deliver outstanding linear properties, enabling it to maintain high precision even at low current levels.
CAT III 1000 V
Key Features
âĒ The ideal current sensor for measuring large 1000 A currents in EV, HEV, and PHEV
âĒ Strengthened shield improves noise resistance to accurately measure current buried in noise
âĒ Flat frequency characteristics and CMRR performance to achieve high accuracy measurement
âĒ More enhanced environmental resistance performance than ever before with operating temperature range of -40 to 85°C
âĒ Basic accuracy of Âą0.04% and phase Âą 0.1°
âĒ Good frequency characteristics over wide band of DC to 1.5 MHz (amplitude) and DC to 1 MHz (phase)
âĒ Ideal for advanced fields such as photovoltaics and fuel cells to evaluate the charge / discharge of batteries and secondary side of inverters
âĒ Connect with oscilloscopes or Hioki Memory HiCorders (using SENSOR UNIT) to monitor waveforms
Model No. (Order Code)
CT6876 | 1000 A AC/DC, ME15W terminal, 3 m (9.84 ft) cable length |
---|---|
CT6876-01 | 1000 A AC/DC, ME15W terminal, 10 m (32.81 ft) cable length |
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Rated current | 1000 A AC/DC |
---|---|
Max. allowable input | Within the derating range, up to ¹1800 Apeak (design value) allowed at 40°C or less for 20 ms or less |
Frequency bandwidth | Amplitude: DC to 1.5 MHz (CT6876), DC to 1.2 MHz (CT6876-01) Phase: DC to 1 MHz |
Basic accuracy | (DC, 45 Hz âĪ f âĪ 66 Hz) Amplitude: Âą0.04 % rdg. Âą0.008 % f.s., Phase: Âą0.1° |
Output voltage rate | 2 mV / A rated (This device outputs AC+DC voltage via the Sensor Unit. Use with voltage input equipment with high input impedance such as 1 MÎĐ) |
Max. rated voltage to earth | 1000 V AC/DC (50/60 Hz, CAT III ) |
Core diameter | Ï 36 mm (1.42 in) |
Operating temperature, humidity | -40°C to +85°C (-40°F to 185°F), 80% RH or less (with no condensation) |
Power supply | Power suppled via the Power Analyzer PW6001, PW3390, or Sensor Unit CT9555, CT9556, CT9557 |
Max. rated power | 7.5 VA max. (at 1000 A/55 Hz) |
Dimensions and mass | 160 mm (6.30 in)W Ã 112 mm (4.41 in)H Ã 50 mm (1.97 in)D, CT6876: 950 g (33.5 oz), cable length 3 m (9.84 ft), CT6876-01: 1250 g (44.1 oz), cable length 10 m (32.81 ft) |
Accessories | Instruction manual Ã1, Mark bands Ã6, Operating precautions Ã1 |
Compatible models
Compatible models | CT6876 |
---|---|
Power Analyzer PW6001, PW3390 | Ok |
Power Analyzer 3390 | Attention (Requires CT9901) CT ratio 2 |
Power HiTester 3193 series | Attention (Requires CT9901) CT ratio 2 |
Current Unit 8971 | Attention (Requires the 9318, CT9901) CT ratio 2 |
Â
Catalog: AC/DC CURRENT SENSOR CT6877, CT6876, CT6875 | Download PDFÂ [2MB] | English |
For more information, click here